Andrew F. Davis 1276187c52 iio: health/afe440x: Make gain settings a modifier for the stages
Currently the TIA gain settings are exported to userspace as sysfs
entries that do not clearly represent their internal relation to the
sampling stages. The gain settings are enabled on a per-stage basis,
this can be seen in figure 24 of the current AFE4404 datasheet.
These gain settings should therefore be tied to the channels that are
read during these stages. Make this change here.

Signed-off-by: Andrew F. Davis <afd@ti.com>
Signed-off-by: Jonathan Cameron <jic23@kernel.org>
2016-05-04 11:11:30 +01:00

42 lines
1.4 KiB
Plaintext

What: /sys/bus/iio/devices/iio:deviceX/in_intensityY_raw
Date: May 2016
KernelVersion:
Contact: Andrew F. Davis <afd@ti.com>
Description:
Get measured values from the ADC for these stages. Y is the
specific stage number corresponding to datasheet stage names
as follows:
1 -> LED2
2 -> ALED2/LED3
3 -> LED1
4 -> ALED1/LED4
Note that channels 5 and 6 represent LED2-ALED2 and LED1-ALED1
respectively which simply helper channels containing the
calculated difference in the value of stage 1 - 2 and 3 - 4.
The values are expressed in 24-bit twos complement.
What: /sys/bus/iio/devices/iio:deviceX/in_intensityY_offset
Date: May 2016
KernelVersion:
Contact: Andrew F. Davis <afd@ti.com>
Description:
Get and set the offset cancellation DAC setting for these
stages. The values are expressed in 5-bit sign-magnitude.
What: /sys/bus/iio/devices/iio:deviceX/in_intensityY_resistance
What: /sys/bus/iio/devices/iio:deviceX/in_intensityY_capacitance
Date: May 2016
KernelVersion:
Contact: Andrew F. Davis <afd@ti.com>
Description:
Get and set the resistance and the capacitance settings for the
Transimpedance Amplifier during the associated stage.
What: /sys/bus/iio/devices/iio:deviceX/out_currentY_raw
Date: May 2016
KernelVersion:
Contact: Andrew F. Davis <afd@ti.com>
Description:
Get and set the LED current for the specified LED active during
this stage. Y is the specific stage number.